The proposed clinical study aims to validate the diagnostic performance, compared to a reference ECG, of the ECG-SW1 software and the PPG-SW1 software integrated into the HWA09 watch developed by Withings for the automatic identification of atrial fibrillation (AF).
Full description
HWA09 is a wrist-worn watch that enables to record a single-lead ECG similar to the standard lead I configuration, and a photoplethysmogram (PPG).
Primary objective :
To validate the diagnostic performance of the ECG-SW1 software integrated into the HWA09 watch for the automatic identification of atrial fibrillation (AF) compared to a simultaneous 12-lead ECG (hereinafter referred to as the "gold standard" or "reference method").
The algorithm classifies the watch signals into four categories: normal sinus rhythm (RSN), AF, arrhythmia other than AF, or non-interpretable. The latter category includes signals of insufficient quality to be interpreted with confidence.
This classification will be compared to the diagnosis made by three independent, blinded cardiologists based on a 12-lead ECG. The final diagnosis retained will be decided by majority with adjudication by a consensus of 3 to 6 cardiologists. The sensitivity and specificity of the device will then be estimated.
First secondary objectives:
The first secondary objective is to validate the diagnostic performance, compared to a simultaneous 12-lead ECG, of the PPG-SW1 software integrated in the HWA09 watch for the automatic identification of atrial fibrillation (AF) from the photoplethysmography (PPG) signal.
Three independent blinded cardiologists will issue a diagnosis based on the ECG recording of the HWA09 watch. The final diagnosis retained will be decided by majority with adjudication by a consensus of 3 to 6 cardiologists. Sensitivity and specificity will be calculated in relation to the reference ECG.
Second secondary objective:
To evaluate the quality of the single-channel strips measured with the HWA09 watch, as follows:
The sponsor will ensure that no cardiologist makes a diagnosis on both a signal from the HWA09 watch and the reference ECG fomr the same patient. The sponsor will centralize diagnostics on the HWA09 and reference ECG signals and perform agreement calculations on the associated pairs.
Enrollment
Sex
Ages
Volunteers
Inclusion criteria
Exclusion criteria
Vulnerable individuals according to the regulation in force :
Individuals having refused to take part in the study
Individuals linguistically or mentally unable to express their non-opposition
Individuals physically unable to wear a watch
Individuals with an electrical pacing by a pacemaker
283 participants in 3 patient groups