Verification of The Cloud DX Vitaliti™ Continuous Vital Sign Monitor (CVSM): Non-invasive Blood Pressure Component

Population Health Research Institute (PHRI) logo

Population Health Research Institute (PHRI)

Status

Completed

Conditions

Blood Pressure

Treatments

Device: Vitaliti

Study type

Interventional

Funder types

Other

Identifiers

NCT03493867
STP-VIT-002

Details and patient eligibility

About

The purpose of this study is the verification of Cloud-DX Vitaliti™ Continuous Automated Non-Invasive Blood Pressure Monitor for Clinical Use and Self Measurement according to standards laid out by IEEE Std 1708 and AAMI-ISO 81060-2.

Full description

Several automated vital sign measurement systems have been developed recently; however, non-invasive continuous blood pressure monitoring continues to be a major limitation. To circumvent the limitations of current systems of interval hemodynamic measurement and increasing demands on health care providers, a non-invasive automated continuous vital sign monitor that can integrate into hospital early warning systems and warn health care providers of deteriorating vital sign parameters would be of significant value and a major advance for current hospital monitoring systems.

Cloud-DX Vitaliti™ is a new low profile automated vital sign monitor, which includes a continuous non-invasive blood pressure measurement system. Additional parameters are also measured. Paired with companion software from cloud based computing, Vitaliti™ features a user friendly interface, onboard radios for communication with smartphones/tablets/personal computers, and several sensors.

Cloud-DX Vitaliti™ has undergone preliminary testing and is now ready for verification to achieve compliance with AAMI/ISO and IEEE standards of medical grade device. At the Hamilton General Hospital (HGH), Cardiac Surgical Intensive Care Unit (CICU) and Vascular Surgical Step Down Units, arterial catheters are standard of care and performed on most patients for variable periods of post-operative recovery. Within this setting, the IEEE Std 1708 and AAMI-ISO 81060-2 would be fully met. Furthermore, the cardiovascular surgical caseloads at HGH would provide sufficient test subjects to validate VitalitiTM in a timely fashion.

Enrollment

78 patients

Sex

All

Ages

18+ years old

Volunteers

No Healthy Volunteers

Inclusion criteria

  1. Patients ≥18 years of age;
  2. Patient has undergone major cardiac or vascular surgery and has been admitted post-op to the cardiac surgical intensive care unit (CICU) or vascular surgery step down unit with an arterial line in place

Exclusion criteria

  1. Patients < 18 years of age;
  2. Pregnant;
  3. Current arrhythmia
  4. Blood pressure outside of test range for phase (refer to table 1)
  5. Allergy to device contact adhesive (e.g. ECG leads)

Trial design

Primary purpose

Other

Allocation

N/A

Interventional model

Single Group Assignment

Masking

None (Open label)

78 participants in 1 patient group

Vitaliti
Other group
Description:
The subject's blood pressure will be simultaneously determined and recorded using the invasive arterial line blood pressure reading and the test Vitaliti device readings.
Treatment:
Device: Vitaliti